AMS 5100 Dewpoint

The basis of the Trace Oxygen Analyser AMS 5100 DP is the development project "AMS 5000" for a compact electronic to be used with Zircon dioxide and electrochemical sensors for the measurement of trace oxygen and Oxygen concentrations in Percent range.

The Trace Oxygen Analyser AMS 5100 DP is suitable for the measurement of oxygen traces in inert gases. For this application the Trace Oxygen Analyser AMS 5100 DP is equipped with a Zircon dioxide sensor. Due to the compact size of the electronics the Trace Oxygen Analyser AMS 5100 DP can be mounted in an electronic housing IP 20, 42 TE, 3 HE and has a depth of 250 mm.

The Trace Oxygen Analyser AMS 5100 DP is available in several housings for General- and Ex-Applications. For determine the dewpoint of the sample gas the analyser coud be equiped with a dewpoint measurement in addition to the oxygen measurement.

AMS 5100

The list of available options includes the following features:

  • electronical/pneumatic gas supply pump
  • electronical flowmeter
  • auto calibration
  • pressure reducer
  • particle filter 2–7µm
  • different housings

Technical data

Analyser AMS 5100 DP
Measuring principle Oxygen: ZrO2 probe with Pt electrodes
Dewpoint: ceramic sensor
Application Gas industries, chemical industries, inert gas applications
Measuring range Oxygen:
0…25 Vol % to 0…1 ppmv
Dewpoint:
– 100 … + 20°C
0...3.000ppm
Analogue signal port (0) 4…20 mA or 0…10V, galvanically separated
one for each measuring component
Reproducability +/- 2 % of the measuring value
Resolution Oxygen:
0,01 ppm – C(O2) – 0,01 %
depending on the O2 concentration
Dewpoint:
0.1 °C
T90-Time approx. 20 seconds for oxygen
approx. 40 seconds for Dewpoint
Display 2* 16 digit, illuminated LCD display
Line 1: oxygen concentration ppm or Vol %
Line 2: dewpoint °C
To adjust settings, you can change the second line of the display.
Messages 1 System message (measuring value yes / no)
1 message each for oxygen / dewpoint
Gas connection inlet / outlet 3 / 6 mm ferrule pack
Gas sampling built-in inlet / outlet valve
Sample flow min. 30 Nl/h, max. 50 Nl/h
sample pressure (inlet) min. 1,01 bar abs., max. 10 bar abs.
Digital communication serial interface RS232
Ambient operating temp. + 5 °C up to + 60 °C
Power supply 110 … 230 VAC or 24 VDC
Protection / Housing / Dimensions IP20 / electronics unit, 42 TE / 3HE
IP65 / wall mounting housing / ca. 300 x 260 x 130 mm (hxbxt)
IP54 / portable housing 42 TE / 3 HE
IP 54 / panel mounting 144 x 144 mm
Ex-d housing for Ex Zone 2
Weight 2,5 kg … 10 kg depending housing and built-in options
Options electronical/pneumatic gas supply pump
electronic flow control
particle filter 2–7µm
auto calibration
max. 4 measuring ranges, automatically switching with digital identification
software for remote control and service
manual 5 way valve
pressure regulator
Version: AMS 5100 Dewpoint V-2013-10