AMS 6000

The AMS 6000 analyser is suitable for measuring traces of oxygen and one other component like humidity, CO, H2 and .. at once. This means it is possible to analyse more than one component at the same measuring point.

For measuring O2, CO2 or H2 an electrochemical sensor is deployed and for the determination of humidity a ceramic sensor is applied. Highly accurate measurement is achievable with the processor-controlled analyser. Changes in concentration of < 0,1 ppm with a resolution of 0,01 ppm can be measured.

AMS 6000

Available options:

  • Pressure reducer
  • Particle filter 2-7 µm
  • Manual 5 way valve
  • Manual purge
  • Electronical/pneumatic gas supply pump
  • Electronical flowmeter
  • Auto-Calibration
  • 2 free adjustable measuring values
  • Different housings
     

Technical data

Analyser AMS 6000
Measuring principle O2, CO, H2 …./ dew point
Application Anwendung
Measuring ranges O2 2 / 4, automatic change-over, digital identification
Measuring ranges CO, H2, dew point 1
Measuring range limit: O2 / CO / H2
min. / max.
Measuring range limit: dew point
adjustable, depends on the sensor
0…10 / 0…10.000 ppm
– 100…+ 20 °C dew point / 0…3.000 ppm humidity
(please specify when ordering)
Analogue Signal Port 2x 0 (4) … 20mA, galvanically isolated, automatic change-over
(one analogue signal port for each measuring component)
Reproducibility +/– 2 % of the measuring value
Resolution (A) O2, CO, H2
Resolution (A) dew point
0,01 ppm < A < 1 ppm depending on measuring component
0,1 °C < A < 1°C / 0,1 ppm < A < 1 ppm
T90 time O2, CO, H2 / dew point 30…40 sec. / 60 sec.
Display 2* 16 digit LCD, back-lit
Messages Alarms, Status, Calibration, Service Request, Floating Change-over Relays
Gas Connection Inlet / Outlet 3 / 6 mm ferrule pack
Gas Sampling Built-in inlet / outlet valve, Flow Meter
Sample Flow min. 20 Nl/h, max. 40 Nl/h
Sample Pressure (inlet) min. 1,01 bar abs., max. 2 bar abs.
Sample Pressure (sensor) max. 50 mbar Excess Pressure
Digital Communication Serial interface RS 232
Ambient Temperature – 5 °C up to + 45 °C
Rel. humidity O2, CO, H2
/ dew point
0 … 99 %
/ please refer to measuring range
Power supply 230 VAC, 24 VDC
Protection / Housing / Dimensions IP 20 electronic housing 19", 3 HE, depth: 473 mm
IP 20 electronic rack ½ 19", 3 HE
Protection class General application
Weight 3 – 7 kg, depends on the components
Options Manual Purge and Bypass Valve
Electronic Flow Monitoring
Ex-Zone 2 applicable
Sample Gas Pump
Particle filter 2-7 µm
Automatic Calibration, remote-control available
Automatic Purging of the Sensor
Pressure Reducer max. 10 bar, out 50 mbar
Version: AMS 6000 E V-2013-07