AMS 8100 / AMS 8150 Ex

The zero air generators AMS 8100 / AMS 8150 are operating on the principle of the catalytic oxydation. At an inlet pressure of up to 15 bar (abs), the AMS 8100 / AMS 8150 generate up to 600 Nl/h of carbonmonoxide- and hydrocarbon free Zero Air. The systems operate highly efficient even at this flow rate; the zero air generators AMS 8100 / AMS 8150 can eliminate 99.95 % of 1000 PPMv CH4 in air.

AMS 8100 / AMS 8150

Technical data

Catalytic air purifier AMS 8100 / AMS 8150 Ex
Ex-classification Ex p IIC T3
Principle catalytic oxydation
Catalyst Pt/Pd on granulated Al2O3
Inlet pressure 15 bar (abs.)
Operating pressure 3 bar (abs.)
Gas flow 600 Nl/h
Time for pre-heating ca. 30 min.
Reactor temperature ca. 400°C
Gas connection 6 mm ferrule pack
Power supply 230 VAC 50/60 Hz
Power consumption 600 VA
Hi / Low Alarm temperature, potential-free relay, 60 V / 1A
Flow monitoring  
AMS 8100 (Option) potential-free relay, 60 V / 1A
AMS 8150 (Standard) potential-free relay, 60 V / 1A
Dimensions HxBxT  
AMS 8100 166 x 520 x 400 mm or 19" rack
AMS 8150 750 x 500 x 300 mm
960 x 500 x 300 mm with filters
Purification of water and nitrogen oxides Molecular sieve
Cleaning the air at the device input (option) Oil separator, Dust filter
Version: AMS 8100 E V-2013-07