AMS 5200 P

The basics of the Oxygen Analyser AMS 5200 is the development project "AMS 5000". The Oxygen Analyser AMS 5200 P is designed for the measurement of Oxygen-mixtures with high O2 concentrations (for example food industry, glass producing industry). For this application the Oxygen Analyser AMS 5200 is equipped with a zirconium dioxide sensor. The main features for the used sensor are the fast response to changes in the gasmatrix and high resistance against pressure fluctuations.

Due to the compact design the Oxygen Analyser AMS 5200 P is available in different housing types. A long list of options allows to adopt the Oxygen Analyser AMS 5200 for almost every application. The software for remote control and remote service with access via the customer computer network further increase the flexibility of the Oxygen Analyser AMS 5200 in any application.

AMS 5200 P

Technical data

Analyser AMS 5200 P
Measuring principle ZrO2 sensor with solid electrolyte
Application Gases Industries, Chemical Industries
Measuring range 0 … 95 Vol %, adjustable
Analogue signal port (0) 4…20 mA or 0…10V, galvanically separated
Reproducability +/- 2 % of the measuring value
Resolution 0,01 – 0,1 %, depending on the measuring range
T90-Time < 20 seconds
Display 2* 16 digit, illuminated LCD display
1. Line: display of concentration in ppm or Vol %
2. Line: Messages, device status, sensor parameters
Messages 1 System message (measuring value yes / no)
Option: 2 messages configurable as oxygen value,
calibration message, measuring value in the range, flow
Gas connection inlet / outlet 3 / 6 mm ferrule pack
Gas sampling built-in inlet / outlet valve
Sample flow min. 30 Nl/h, max. 50 Nl/h
sample pressure (inlet) min. 1,5 bar abs., max. 5 bar abs.
Digital communication serial interface RS232
Ambient operating temp. + 20 °C up to + 60 °C
Power supply 110 … 230 VAC or 24 VDC
Protection / Housing / Dimensions IP20 / electronics unit, 1/2 inch, 19 inch / 3HU
IP65 / wall mounting housing / ca. 300 x 260 x 130 mm (hxbxt)
IP54 / portable housing 42 TE / 3 HU
Weight 2,5 kg … 6 kg depending housing and built-in options
Options Software for remote display and service
manual 5 way valve
Particle filter 2-7µm
max. 4 measuring ranges automatically switching with digital range detection
pressure regulator
Version: AMS 5200 P E V-2019-06