TCD 200 TG for binary gas mixtures

For measurement of binary gas mixtures AMS offers the Thermal Conductivity Analyser TCD 200 TG.

The measuring principle of Thermal Conductivity uses the different thermal conductivity of gases. The gas analysis through thermal conductivity for the measurement of gas components is a common method since the early 20th. century.

Typical applications are the measurement of:

  • Hydrogen in Nitrogen
  • Hydrogen in Helium
  • Argon in Nitrogen or Oxygen 

The main advantage of the Thermal Conductivity Analyser TCD 200 TG is the speed of the measurement. Due to the extremely low thermal mass of the sensor, T-90 Times of less than 10 Seconds at a gas flow of 50 Nl/H can be reached.

Typical gas mixtures for the use of the Thermal Conductivity Analyser TCD 200 TG are:

Sample gasAssociated gasMeasuring range
H2N2 or air0 … 0,5 Vol. %
H2N2 or air0 … 100 Vol. %
H2N2 or air98 … 100 Vol. %
H2Ar0 … 0,3 Vol. %
H2He20 … 100 Vol. %
HeN2 or air0 … 0,8 Vol. %
HeN2 or air0 … 100 Vol. %
HeN2 or air97 … 100 Vol. %
ArN2, O2 or air0 … 3 Vol. %
ArN2, O2 or air96 … 100 Vol. %
ArN2, O2 or air0 … 100 Vol. %
CH4N2 or air0 … 3 Vol. %
CH4N2 or air96 … 100 Vol. %
CH4N2 or air0 … 100 Vol. %

Technical data

Measuring systemTCD 200 TG
Measuring principleThermal Conductivity
Measuring rangeDepending on gas composition
i. e. H2 / N2, CO2 / N2, CH4 / N2, Ar / N2
Analogue output port0 … 10 V, (0) 4 … 20mA
Reproducability< 1 % of measuring range
Resolution0,1 % of measuring range, or 0,01 Vol. %
depending on gas concentration
T-90 Time< 10 Sec. at a flow of 50 Nl / h
Longterm drift< 1 % of the smallest measuring range per month
Display2 x 16 digit LCD Display
Messages2 freely programmable messages with reference to the gas concentration
GasanschlussEingang / Ausgang 3 / 6 mm Klemmringverschraubung
Gas samplingintegrated gas inlet valve
Gas flow meter
Sample gas flowmin. 25 Nl/h, max. 50 Nl/h
Sample gas pressuremin. 1,01 bar abs., max. 2 bar abs.
Sample gas temperature5 °C bis + 45 °C
Ambient temperature5 °C bis + 45 °C
Relative humidity0 … 99 % not condensing
Power supply24 VDC
Protection class / Housing /IP 20 / electronics unit, 28 TE / 3HU
DimentsionsIP 65 / wall mounting housing / ca. 300 x 260 x 130 mm (hxbxt)
IP 54 / portable housing 42 TE / 3 HU
IP 54 / panel mounting 144 x 144 mm
Weight2-5 kg
Optionsinert gas purging for preventing hazardous atmosphere inside housing
pressure reducer (max 10 bar in, 50 mbar out)
particle filter 2 µm
5 way valve
electrical/pneumatical pump
electronic flow monitoring /-alarm
Version: TCD200TG-E V-2013-07